|
The hotwire anemometer accompanied by a feedback circuit is referred to as the
constant temperature anemometer(CTA), Figure 3.17 - 3.18. The CTA consists of a Wheatstone
bridge and a servo amplifier. One arm of the Wheatstone bridge is the probe
sensor. As the flow condition varies, the sensor tends to cool appropriately with a resulting
change in resistance.The change in resistance leads to an error voltage .These two voltages form the input to the operational amplifier. The selected amplifier
has an output
current, which is inversely proportional to the change in the resistance
of the hot-wire sensor. Feeding
this current back to the top of the bridge will restore the sensor's resistance to its original value. The feedback circuit plays an important role in
improving the frequency response of the hotwire, typically from 10-20 Hz to several kHz.
The temporal flow feild has been mapped in the present work using an X-probe
for obtaining the u and v
components of velocity. In two-dimensional measurements,
calculation of velocity components involves
solving a pair of non-linear, simultaneous
equations. Thus, the accuracy of hotwire measurements is affected by the accuracy of
the calibration procedure. In addition, the accuracy of the numerical solution
that is
used to solve the nonlinear simultaneous equations is relevant. Two approaches were
applied
for data reduction and are discussed in Chapter 15. Both approaches yielded
nearly identical velocities,
thus showing that numerical aspect of data reduction was
satisfactory.
The X-probe supplied by DANTEC has been used in the present research, which
mostly satisfies
all the above- stated criteria. Additionally, some of recommended electronic tests by the manufacturer
have also been carried out to optimize the response
of the anemometer's output voltage.
|